Propulse cleaning tablets
Propulse cleaning tablets for thorough cleaning of the entire ProPulse system. Very simple dosing so that the device can be cleaned effectively and no damage can occur to the ProPulse if these tablets are used correctly.
Cleaning instruction
- Place a Propulse cleaning tablet in the reservoir, fill the reservoir with water up to the 500 ml mark and wait until the tablet is completely dissolved.
- Turn on the Propulse NG Electronic Ear Irrigation System for a few seconds to prime the pump and tubing.
- Now leave it like this for 10 minutes.
- Remove the solution from the reservoir, then refill the reservoir with chilled boiled water or cold tap water and rinse the entire system so that no cleaning solvent remains.
- This cleaning procedure should be performed at the beginning of each day the earwash unit is used.
- Do not leave the solution longer than stated in the unit.
Delivery
- In a resealable jar
- 200 chlorine cleaning tablets
To view this video, you first have to accept the cookies.