Coagulator Diatermo MB 160 - mono-bipolar 160W
High frequency electrosurgical unit suitable for monopolar or bipolar electrosurgery.
Suitable for cutting (CUT), cut coagulation (BLEND), superficial coagulation (FORCED COAG), deep coagulation in absence of necrosis (SOFT COAG) and with forceps, bipolar coagulation.
The use of distributed or normal neutral plate is possible.
Applications:
MB 160 is recommended in the case of humid/moist operating conditions, for example, mucus, or with a consistent necessity to have a reserve of usable power, therefore, the MB 160 suitable for gynecological and orthopedic surgery.
Leading Edge Technology, this advanced concept allows the device to be used for very professional operational procedures thanks to ergonomic and safety features.
Open Circuit (OC) Alarm warns the user when the connection with the neutral plate isn't made, or if the cable is disconnected or the neutral plate is not connected with the patient. The alarm is shown by an intermittent beep together with a flashing red light on the front panel. In these cases, the high frequency for cutting and coagulation is disabled.
OVT Alarm, the unit is equipped with an internal circuit protection warning when the current operation is taking too long.
Technical specifications:
- Minimal pre-selectable power: 0
- Level step: 1
- Maximum output power CUT (W): 120-250 ohm
- Maximum output power BLEND (W): 90-200 ohm
- Maximum output power FORCED COAG (W): 80-150 ohm
- Maximum output COAC SOFT (W): 60-100 ohm
- Maximum output power BIPOLAR (W): 40-100 ohm
- Crest Factor CUT: 1.5
- Crest Factor BLEND: 2.1
- Crest Factor COAG FORCED: 2.0
- Crest Factor COAG SOFT: 1.7
- Crest Factor BIPOLAR: 1.5
- Frequency: 600 kHz
- Weight: 7 kg
- LxHxD mm Size: 260 x 110 x 265
- Voltage (switchable): 115-230
- Frequency (Hz): 50-60
- Fuses 230Vac (5x20) Timed: 2x 3.15A
- Fuses 115Vac (5x20) Timed: 2x 6.3A
- Electrical input power (VA): 300
- Electrical input current (230Vac) (A): 1.3
- Electrical input current (115Vac) (A): 2.6
- Five steps adjustable sound level (from 55 to 75 dBA)
- Self-check monitoring.
- Power alert.
- Divided or normal plate.
- Remembers the last used settings.
- Electrical Class (EN60601 - 1) I CF
- MDD 93/42 / EEC Class: IIB
- Patient circuit: F
- Conforms to EN60601-1-4 (1998)
Standard accessories:
- Handle with two buttons single use.
- MB foot switch (to be used as an alternative to the grip)
- MB flat neutral plate and cable
- Kit 10 short autoclavable electrodes, Ø 2,4 mm
- Power cable 2 m
- User Manual: English, Spanish, Italian, French, Polish
To view this video, you first have to accept the cookies.